Instructions: Filter pressure
alarm
Step
1
2
Instructions: TMP alarm
Step
1
2
UniFlux 10 Instrucciones de funcionamiento 28993587 AE
Action
In the Text pane, select the Base Time instruction of the Block
INITIAL_CONDITIONS block.
In the Instruction box, do the following settings:
Set Breakpoint to 0.00 min.
•
In the Instructions group box, select Alarms and DeltaP_Alarms.
•
In the Parameters field, set High Alarm, Low Alarm, High Warn and
•
Low Warn to appropriate values according to the specifications for the
filter in use.
In the Parameters field, set Mode to Enabled.
•
Click the Insert button.
•
Action
In the Text pane, select the Base Time instruction of the Block
INITIAL_CONDITIONS block.
In the Instruction box, do the following settings:
Set Breakpoint to 0.00 min.
•
In the Instructions group box, select Alarms and TMP_Alarms.
•
In the Parameters field, set High Alarm, Low Alarm, High Warn and
•
Low Warn to appropriate values according to the specifications for the
filter in use or requirements of the intended application (if lower than
filter specifications).
In the Parameters field, set Mode to Enabled.
•
Click the Insert button.
•
A Create a method with UNICORN 5
A.2 An example of using the Method Editor
A.2.3 Block 1: Initial conditions
199