A Create a method with UNICORN 5
A.2 An example of using the Method Editor
A.2.4 Block 2: Product transfer
Instruction: Tank level watch
Step
1
2
End block: Allow the tank to be
filled
Step
1
2
204
Action
In the Text pane, select the Base Time instruction of the Block
PRODUCT_TRANSFER_VOLUME block.
In the Instruction box:
Set Breakpoint to 0.00 min.
In the Instructions group box, select Watch and WatchTankLevel.
Set:
1 Test to Greater_Than.
2 Value to a number in liters given by the tank specification or the
application requirements.
3 Action to END_BLOCK.
Click the Insert button.
Action
In the Text pane, select the End_Block instruction of the Block
PRODUCT_TRANSFER_VOLUME block.
In the Instruction box:
Set Breakpoint to a value high enough for the tank to reach the volume
set for WatchTankLevel (in the previous step).
Click the Change button.