ResMed S8 Compact System Manual De Instrucciones página 32

Ocultar thumbs Ver también para S8 Compact System:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

338249r1.book Page 27 Friday, June 10, 2005 3:22 PM
Recommended separation distances between portable and mobile RF communications
equipment and the S8 series of devices
The S8 series of devices is intended for use in an environment in which radiated RF disturbances are controlled.
The customer or the user of the S8 device can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the S8 device
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
150kHz to 80MHz
80 MHz to 800 MHz
d = 1.17 √P
d = 0.35 √P
0.17
0.37
1.17
3.69
11.70
m
800MHz to 2.5 GHz
d = 0.7 √P
0.04
0.07
0.11
0.22
0.35
0.7
1.11
2.21
3.50
7.0
S
S
YSTEM
PECIFICATIONS
27
Tabla de contenido

Solución de problemas

loading

Tabla de contenido