polymerization, pressing this button again allows the
current cycle to be interrupted at any time.
Exposure time
Exposure time is 10 seconds. During polymerization, a
microflash warns when the cycle is halfway through.
Indicator light (4)
The handpiece comprises a two-color indicator light
located above the on/off button.
- During normal operation, when a polymerization cycle
is run, the green indicator light is lit
- When the device detects a low charge condition, the
indicator shines a steady red.
Technical specifications
Model name:
Medical category:
Dimensions without light guide:
Weight:
Dimensions:
Operation:
Protection:
Category:
Protection:
Battery inverse polarity protection
Protection index:
Batteries (not supplied)
Type:
Quantity:
Optical specifications:
- LED for polymerization:
Wavelength range:
Central wavelength:
Intensity:
light guide Ø7,5mm (Ø active 6.8mm).
Maximum exposure time:
Temperatures:
Operation:
Storage:
Humidity:
Operation:
Mini LED Black
I
135g
(without batteries)
Ø25 x 210mm
Continuous operation
Type B
3AT FU1 fuse (non
accessible) 125 V
IPX0
AA
3
420nm - 480nm
455nm – 465nm
1250mW/cm² ± 10% with
10 seconds (show 15)
+10°C to +30°C
-20°C to +70°C
30% to 75%
Storage:
Atmospheric pressure
Autonomy (for information purposes only):
Alkaline batteries: 30 to 100 cycles according to battery
quality
Ni-Mh batteries: 170 to 500 cycles according to capacity.
IV – STARTUP
4.1 UNPACKING THE UNIT
On reception of the unit, look for any damage that may
have occurred during transportation.
If necessary, contact your supplier.
4.2 RECOMMENDATIONS
Check that the environmental conditions have been
complied with.
4.3 INSTALLATION
- Remove the protective cover from the handpiece
- Unscrew the rear cover of the battery compartment
- Insert the batteries/rechargeable batteries making
sure the polarity mentioned on the handpiece is
correct.
- Tighten the screw on the rear cover.
The indicator light blinks red briefly then changes to
green which indicates a good charge level for the
batteries.
V – ROUTINE OPERATION
After inserting the optical guide, insert the flexible cup
or the rigid protection shield.
Press the on/off button to start the polymerization
cycle.
A 10-second polymerization cycle then begins. After 5
seconds, a microflash alerts that half of the cycle has
been reached.
During polymerization, pressing the on/off button again
allows the current cycle to be interrupted.
VI – MAINTENANCE
Before cleaning the handpiece, insert the protective
cap in the place of the light guide, supplied to ensure
that liquid does not enter the handpiece.
7
10% to 100%
condensation included
700hPa to 1060hPa