Working And Use; Working; Reliability Of The Test Validity - Tecno-gaz SEE-TEST 260/S Istrucciones Para El Uso

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Par 05
Working and application
SEE-TEST is a device for medical and/or dental use.
TECNO-GAZ will not be held responsible for damage to the machine or harm to people caused by
the improper use of the machine or by the user's unskilfulness.
ROOM TEMPERATURE +10 / +30 °C
5.1

Working

1. Select the temperature cycle on the temperature selector switch (POS. E – PICT. 2A)
Position 0 corresponds to temperature cycle 37°C
Position 1 corresponds to temperature cycle 57°C
2.
Select position 1 on the general switch (POS. D – PICT. 2A), push START button (POS. G – PICT.
2C): "ON" and "HEATER" led lighten. "HEATER" led keep lighten till reaching the selected
temperature, then "°C" led lighten and a sound signal advise of the reaching of the selected
temperature. "°C" led also means that the selected temperature keep even. When the temperature
decrease, due to the natural cooling of the heating element, the resistance begins to work. "ON" led
always lighten, this means that the device is working.
5.2

Reliability of the test validity

As already said, "ON" led means the device regular working, in case of lack of electrical energy, or broken
control system, the device switch off permanently and led keep turned off, also after the reset of the
electrical energy. In this way the user, incubation time past, can verify, if led are switched on, that the test
is valid. Otherwise, led switch off means an electrical energy interruption or a broken control system. The
test is not valid.
31
ENGLISH
DSTF080_REV5
Tabla de contenido
loading

Tabla de contenido