Review Of The Technical Status; Mechanical Faults; Recurring Safety Inspections; Servicing And Manufacturer Address - HEBU medical HBS x-touch 300 BiSeal Instrucciones Para El Uso

Aparato de cirugía electrodoméstica
Ocultar thumbs Ver también para HBS x-touch 300 BiSeal:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

EN
Electromagnetic safety
In the development of electrosurgical units, HEBUmedical attaches particular importance to the
stringent electromagnetic emission regulations. Consequently, the solutions we supply have a
correspondingly minimal emission level for the required output. Measurements conform the high level
of electromagnetic safety offered by HEBUmedical electrosurgical units.
Under typical working conditions based on an 8-hour day, a field occurs at a distance of 5 to 15 cm
from the cables. At a distance of 20-40 cm, the value drops well below the prescribed limit.
Electromagnetic fields occur primarily around the cable.
The electrosurgical unit itself does not constitute a significant emitting element.

10. Review of the technical status

Before any procedure, check the cables, accessories (handles, electrodes, forceps) and foot switches
for any signs of visible irregularity. Do not use cables or instruments with broken or defective
insulation. These can result in burns if they come into contact with the skin!

10.1 Mechanical faults

If damage occurs to switches, connecting sockets, houses or the touch screen, or if the unit is
dropped on the floor, please contact your authorized dealer for advice on further use.
The dealer can arrange for a detailed technical check to be performed.
The manufacturer accepts no liability for calibration work or repairs to the unit carried out by the user.

10.2 Recurring safety inspections

A safety inspection of the electrosurgical system is required once a year. The manufacturer only
permits the use of units for which an up-to-date inspection by suitably trained personnel can be
verified.
All information relating to the inspections (date, inspecting staff member, result etc.) must be recorded
in the unit's log book.

10.3 Servicing and manufacturer address

HEBUmedical GmbH
Badstraße 8
78532 Tuttlingen / Germany
Tel.
+49 7461 94 71 - 0
Fax
+49 7461 94 71 - 22
Web: www.HEBUmedical.de
54
®
HBS x-touch 300 BiSeal
Tabla de contenido
loading

Este manual también es adecuado para:

Hf 9509

Tabla de contenido