en
MAXN
Neonatal/Adult SpO
Identification of a substance that
is contained or present within the
product or packaging.
Do not use if
package is opened or
damaged
This product cannot be adequately cleaned and/or sterilized by the
user in order to facilitate safe reuse, and is therefore intended for single
use. Attempts to clean or sterilize these devices may result in a bio-
incompatibility, infection or product failure risks to the patient.
This product contains DEHP. The intended use limits exposure to transient
contact, minimizing the risk of DEHP release from the device. In order to
avoid undue risk of DEHP exposure in children and nursing or pregnant
women, product should only be used as directed.
Directions for Use
Indications/Contraindications
The Nellcor™ Neonatal-Adult SpO
single patient use when continuous noninvasive arterial oxygen saturation
and pulse rate monitoring are required for neonates weighing less than
3 kg or adults weighing more than 40 kg.
The MAXN is contraindicated for use on patients who exhibit allergic
reactions to the adhesive tape.
Note: Adhesive sensor consists of sensor, cable and wrap adhesive flaps.
Instructions for Use
1. Remove plastic backing from the MAXN and locate transparent
windows on the adhesive side. Windows cover optical components.
Note corresponding alignment marks (a) on nonadhesive side and
dashed line (b) midway between the marks. (1)
2. Orient the MAXN so the dashed line is on the lateral edge of the site.
Neonates: The preferred site is a foot. Alternatively, use a hand. The
window next to the cable goes on the sole of the foot as shown. (2)
Adults: The preferred site is an index finger. Alternatively, other fingers
may be used. The window next to the cable goes on the nail side, distal
to the first joint. Do not place on a joint. Note that the cable must be
positioned on the top of the hand. (3)
Note: When selecting a sensor site, priority should be given to an
extremity free of an arterial catheter, blood pressure cuff, or
intravascular infusion line.
Sensor
2
Sensor, model MAXN, is indicated for
2
Identification of a substance that is
not contained or present within the
product or packaging.
Protected against access to hazardous
22
parts with a finger and protected
against vertically falling water
drops when sensor tilted up to 15°.
3 3