Pelton & Crane Validator Plus Instrucciones De Uso página 40

Tabla de contenido
CHECK
COMPONENTS
DISPLAY
CHECKED
Fill/vent solenoid
Dump solenoid
Bellows solenoid
Steam sensor
Surface sensor
Battery
Pressure trans-
ducer
Line voltage
OPERATIONAL
FUNCTION
CHECKED
Solenoid operation and
electrical continuity.
Solenoid operation and
electrical continuity.
Solenoid operation and
electrical continuity.
Electrical continuity and
parameter check.
Electrical continuity.
Parameter memory for
specials mode (e.g.,
clock settings).
Nominal operating
parameter.
Adequate voltage.
-40-
MAINTENANCE
Self-Diagnostic Check
Automatic Check
The autoclave automatically performs a diagnostic self-check
on nine key components for operational continuity each time
the main power switch (toggle switch located at rear of unit) is
turned on.
An additional check may be made by the following procedure:
Operational Check
Depress "Power On"
switch and "Clear/Start"
the same time. The autoclave beeps once, signaling the
beginning of the self-diagnostic procedure.
During the first four seconds, check that all display segments
illuminate. If segments do not illuminate, call an authorized
service representative. This problem can give the operator
incorrect information if not repaired.
After self-diagnostics, the unit returns to stand-by. The lower
window displays any error found during the Operational or
Automatic self-diagnostic check procedure. See table at left
for explanation of display.
Call an authorized service representative if an unsatisfactory
check is indicated.
switch at
Tabla de contenido
loading

Tabla de contenido