Technical Data - ProMinent DULCOTEST CLT 1 Serie Instrucciones De Servicio

Medidor de clorito
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

14 Technical Data

Measured
variables
Chlorite (ClO
Area of
application
Potable water
Measuring range
CLT 1-mA-2 ppm:
CLT 1-mA-0.5 ppm: 0.02 - 0.5 mg/l
Resolution
CLT 1-mA-2 ppm:
CLT 1-mA-0.5 ppm: 0.01 ppm
Standard slope
of signal
at pH 7.5, 15 °C, Flow = 50 l/h in DGMa
CLT 1-mA-2 ppm:
CLT 1-mA-0.5 ppm:
Response time
T
pH-range
6.5 - 9.5
Conductivity
water sample
0.05 - 5 mS/cm
Temperature
range
1-40 °C, temperature compensation integrated, no sudden rises in tempera-
ture allowed
Speed of tem-
perature change
max 0.3 °C/min
Pressure
Sample water in in-line housing:
Flow
Sample water through DLG III and DGM in-line
probe housing:
Cross sensitivity
Fe
Service life
membrane cap
Typically 3 - 6 months, depending on water quality
Materials
Membrane cap:
Electrode shaft:
Supply voltage
16 - 24 V DC; min 35 mA at 16 V DC
Output signal
4 - 20 mA
Enclosure rating
IP 65
Storage
temperature
between 5 °C and 50 °C
®
ProMinent
)
2
0.1 - 2 mg/l
0.02 ppm
= approx. 60 s (where concentration is rising or falling)
90
2+
, Mn
2+
, nitrite, ... and other oxidising substances
PPE
PVC
6 mA/ppm
24 mA/ppm
max. 1 bar (no negative pressure
and no peaks in pressure permitted)
optimum:
50 l/h
minimum:
20 l/h
maximum: 100 l/h
Technical Data
43
Tabla de contenido
loading

Este manual también es adecuado para:

Dulcotest clt 1-ma-2 ppmDulcotest clt 1-ma-0,5 ppm