Application example
Measuring
transducer
Fluoride sensor
1 Supply unit for two-wire measuring transducer with electrical insulation.
2 If several inputs are present on the unit they must be electrically insulated.

Technical data

Measuring range:
Accuracy:
Signal output:
Input resistance:
Connecting thread:
Power supply:
Maximum power
consumption:
Ambient temperature: -5...+50 °C,
Protection type:
Material:
Dimensions:
Weight:
Part No:
Accessories
2-wire sensor cable:
Reference
electrode
0 - 213.8 mV ^ = 0.03 - 130 ppm with measuring arrangement consisting
of fluoride sensor FLE 010 SE and reference electrode REFP SE
≥ ± 1 mV
4...20 mA ^ = 0...213.8 mV
Output signal is uncalibrated.
Signal output is not electrically insulated from signal input.
11
5 x 10
SN6
24 V DC (18-28 V DC)
0.5 W
maximum 90 % relative humidity
IP 65
PPE
138 x Ø 25 mm
85 g
1009962
External diameter:
Conductor cross-section:
Conductor resistance:
Capacitance conduct./conduct.: 90 pF/m
=
Supply unit
SPS
µP-indicator
Controller
PC
4 mm
2 x 0.25 mm
max. 80 Ω/km
1
~
2
Part No. 725122
2