Functional Description
EOI
÷Err
or
HI
-Err
or
LO
LO BAT

1.3 User Information

Light absorption too great. Reason: e.g. dirty optics
Measuring range exceeded or excessive turbidity
Result below lowest limit of measuring range
Replace 9 V battery, no further analysis possible

1.4 Safety Information

- The reference reagent is corrosive. Wear corresponding
protective clothing (goggles/gloves).
- Request safety datasheets as required.
- Dispose of reagent solution correctly.

1.5 Technical data

Light source:
Battery:
Auto-OFF:
Ambient conditions:
CE:
15
LED, Filter (λ = 430 nm)
LED, Filter (λ = 528 nm)
9 V-block battery
(Life 600 tests)
Automatic switch-off 5 minutes after last
keypress
5-40°C
30-90% rel. humidity (non-condensing)
DIN EN 55 022, 61 000-4-2,
61 000-4-8, 50 082-2, 50 081-1,
DIN V ENV 50 140, 50 204