SimplyGo Mini User Manual
Visual, Audio
Indicators
1 beep
that repeats every
16 seconds
34
Description
Low Oxygen
Concentration Alarm
This alarm occurs when
the device is delivering
a lower concentration of
oxygen than specified�
Note: This alarm occurs
when the internal O
sensor reads <82% O
will turn off if the system
is able to recover from the
fault and achieve >84% O
Press the yellow Alarm
silence button on the
screen to temporarily
silence the alarm, and
the 2nd screen shown
at left appears� Or, press
the Home button on the
device to return to the
Home screen, and the 3rd
screen will appear with
the Low O
alarm symbol
2
in the upper left corner to
indicate the alarm state�
If in normal operation, the
internal oxygen sensor
monitors O
purity during
2
the warm-up period and
then at 1 hour intervals
after that�
Note: This alarm can
take up to 61 minutes to
activate from the time the
alarm condition is present�
What To Do
2
� It
2
2
Change to another
source of oxygen
and contact your
equipment provider�