Caring For Your Simplygo Mini System; Cleaning The Device And Rechargeable Battery - Philips Respironics SimplyGo Mini Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Storage at temperatures above 113° F (45° C), such as in a hot car, may
degrade battery performance and reduce battery life� Storage at low
temperatures may affect initial battery performance�
The optimal temperature range for charging the batteries is 0° F to 113° F (0° C
to 45° C) at up to 80% relative humidity� The device will not charge the battery
if the battery cells exceed 113° F (45° C)� Battery charging will be inhibited
until the battery temperature returns to 111° F (44° C) or below�

Caring for your SimplyGo Mini System

Cleaning the Device and Rechargeable Battery

Warning: To avoid electrical shock, do not remove the SimplyGo Mini
covers� The covers should only be removed by authorized service
personnel� Do not apply liquid directly to the covers� Do not use
alcohol, solvents, polishes, or any oily substances on the device, as
they are flammable�
Caution: Do not allow liquids into any of the controls, the interior of the
case, or the oxygen tubing connector� If this occurs, contact your
equipment provider for assistance�
Cleaning
The exterior covers of the device and rechargeable battery should be cleaned
between patient use and as needed by performing the following steps:
1�
Turn the device off and disconnect from the power source before
cleaning�
2�
Remove the device from the carrying case�
3�
If the battery is removed, wipe the battery bay and battery with a dry
cloth�
4�
Clean the device exterior using a damp cloth with a mild household
cleaner and wipe it dry�
Disinfection
The exterior covers of the device and rechargeable battery should be
disinfected between patient use as follows:
1�
Clean the device as indicated above�
SimplyGo Mini User Manual
43
Tabla de contenido

Solución de problemas

loading

Tabla de contenido