15.2 Alarm Overview and Troubleshooting
Alarm Message
Alarm Cause
The programmed total volume
has been completely given.
END
The target volume programmed is
less than or equal to than the
END OF THERAPY
VOLUME GIVEN THIS
500/500 mL
NUTRITIONAL THERAPY
The pump door is open or the door
mechanism is damaged.
DOOR OPEN
Administration set inserted
incorrectly or not at all.
NO CASSETTE
Feed container and administration
set are empty down to the pump.
Administration set not adequately
AIR IN LINE
filled.
Solution
End administration or program a new nutritional therapy.
Clear
VOLUME GIVEN THIS THERAPY
The reprogrammed target volume must be higher than the
.
THIS THERAPY
Note:
If not cleared before each new nutritional therapy, the end of therapy alarm will
.
sound prematurely.
Close the pump door.
If damaged, contact your local Nestlé Health Science representative or Nestlé
approved service provider.
Open the door and check for the correct installation of the cassette and tubing.
Refer to page 2 and section 6.4 of the user manual.
If nutritional therapy is finished, end the administration, switch pump off and flush
feeding tube.
If nutritional therapy is not finished:
Replace or refill tube feeding container, prime the administration set and restart
feeding or,
Check for and release the occlusion, if any, between the container and the pump and
restart feeding or,
Ensure the administration set is completely filled and prime if necessary and restart
feeding.
Precaution:
Ensure that the administration set is not connected to a patient
when priming.
to start a new nutritional therapy.
VOLUME GIVEN