Button Function; Display - Nestle Health Science Compat ella Instrucciones De Uso

Bomba portátil de nutrición enteral
Ocultar thumbs Ver también para Compat ella:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 61

3.3 Button Function

The pump always retains in memory the on-screen settings and values when scrolling from one screen to the next, as well as when the pump is turned off.
Button
Primary Function
Turn pump on or off (hold 2 seconds (2"))
Start / Pause pump
Auto-prime (hold 2 seconds (2"))
Press again to stop auto-prime early
Scroll through rate, volume and interval programming
Decrease and increase values (hold: accelerates speed)
Clear the on-screen value to default or 0
Scroll through recent feed history
Scroll through pump settings
Precaution:
Ensure that the administration set is not connected to a patient when priming.

3.4 Display

Warning:
During pump start up, confirm all the screen segments shown in the diagram hereafter are visible. If any are missing or if the screen is not
readable, do not use the pump.
Screen Layout
The Compat Ella
pump features a colour display with alphanumeric characters, symbols and a backlight.
®
Secondary Function
Snooze alarm
Switch between continuous and Interval modes
(hold 5 seconds (5"))
Lock and unlock keypad (press both buttons together)
Reset all program values to default and clear recent feed
history (hold button 3 seconds (3"))
Access last rolling 30-day Volume History
(hold button 2 seconds (2"))
Change language (See section 12.2)
EN
37
Tabla de contenido
loading

Tabla de contenido