Guidance and manufacturer's declaration – electromagnetic immunity
The Compat Ella
pump is intended for use in the electromagnetic environment specified below.
®
The customer or the user of the Compat Ella
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic envi-
ronment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location
in which the Compat Ella
pump is used exceeds the applicable RF compliance level above, the Compat Ella
®
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or reloca-
ting the Compat Ella
pump.
®
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
pump should assure that it is used in such an environment.
®
Compliance level
Electromagnetic environment - guidance
3 Vrms
Portable and mobile RF communications equipment should be used no
closer to any part of the Compat Ella
commended separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
3 V/m
d=1.2√P
d=1.2√P 80 MHz to 800 MHz
d=2.3√P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electro-
magnetic site survey,
frequency range
marked with the following symbol:
pump, including cables, than the re-
®
should be less than the compliance level in each
a
. Interference may occur in the vicinity of equipment
b
pump should be observed
®