4.1 Continuous Mode: Menu Overview
Press
to scroll through
SET RATE
Pressing
again cycles back to
Button Device Display Function
The delivery speed in mL / hr at which
the pump will run.
• Rate may be set from 1-600 mL / hr in
SET RATE
• Default value 0 mL /hr
The volume of tube feed to be given, in
mL.
SET
• Volume may be set from 1-4000 mL in
VOLUME
• D efault value:
Return to
SET RATE
Note:
•
Both
and
RATE
VOLUME
changed or the pump has just been turned on. Only then can the pump
be started.
* Continuous: no feeding volume has been set. The pump will run conti-
nously until the formula runs out.
and
.
SET VOLUME
SET RATE
1 mL / hr increments.
1 mL increments.
*
CONT
screen
SET RATE
must be set / reviewed if either one is
4.2 Interval Mode: Menu Overview
Press
to scroll through
SET RATE
and
SET INTERVAL
SET TOTAL VOLUME
cycles back to
SET RATE
Button Device Display Function
The delivery speed in mL/hr at which
the pump will run.
SET
• Rate may be set from 1-600 mL/hr in
RATE
• Default: 0 mL/hr
The bolus volume of tube feed to be
given before each interval pause, in mL.
SET
• Bolus volumes may be set from 1-2000
BOLUS
• Default: 0 mL
The interval between start of one bolus
and start of the next bolus.
• Interval may be set from 00:15-12:00
SET
INTERVAL
• Default: 00:15 hr
The total volume in mL to be given
during the entire nutritional therapy.
SET
• Total volume may be set from 1-4000
TOTAL
VOLUME
• Default value:
SET
Return to
RATE
and
,
SET BOLUS
. Pressing
1 mL/hr increments.
mL in 1mL increments.
hr in 15-minute increments. Automati-
cally calculated when Rate and Bolus
have been programmed.
mL in 1mL increments.
*
CONT
screen
SET RATE
EN
41
again