6.3
Power On and Off
Button
Direction / Result
Power On
Press and hold
systems and will display all screen segments briefly as well as the time. Ensure that all the segments are working and
that the time is correct.
Power Off
Press and hold
2"
Note:
the pump is next turned on.
Warning:
When turning the pump on, if you do not hear the tones or do not see all the screen segments, do not use the pump. Contact your Nestlé Health
Science representative or Nestlé approved service provider.
Power Status Light
Solid green = Pump is plugged into the AC power.
6.4 Inserting the Administration Set
Warnings:
•
Use only Compat Ella
•
Inspect the Compat Ella
•
Replace the Compat Ella
Loading the Cassette
See sequence on page 2 to load set into the pump.
Fig. 1:
open door by lifting the thumb tab
Fig. 2, 3: wrap cassette loop around rotor. Note the arrow which shows correct orientation
Fig. 4, 5: using your thumb, slide the cassette to the right to clip the cassette in the pump
Fig. 6:
close the door by pressing on the "PUSH" symbol.
Note:
Compat Ella
Administration Sets are equipped with a patented Auto-Anti-Free-Flow (AAFF) mechanism, which is always active when needed.
®
The AAFF mechanism is only released when the cassette is inserted in the pump and the door is closed.
for 2 seconds (2") to power on. You should hear 3 tones. The pump will perform a diagnostic of all
2 seconds (2") to power off. You should hear 3 tones as the pump shuts down.
All nutritional therapy parameters, volume history and user settings are retained in the pump's memory for re-use when
administration sets.
®
administration set. Do not use if damaged or altered.
®
administration set every 24 hours to maintain delivery accuracy and help prevent the growth of potential harmful contamination.
®