Intended use
Single-stage fixed pressure reducer without equali-
zation of the input pressure, to reduce the pressure
of compressed medical gases.
To be connected to compressed gas cylinders
Reduction from a maximum input pressure of
200 kPa x 100 (200 bar or 2900 psi)* to an out-
put pressure of 5 kPa x 100 (5 bar or 72.5 psi)
Available for medical compressed air (AIR), oxy-
gen (O
), carbon dioxide (CO
)** or laughing gas
2
2
(N
O), see gas type specification on the pressure
2
reducer
For supplying suitable medical devices in accord-
ance with the technical data on the rating plate
and in the instructions for use
For stationary and mobile use in rooms used for
medical purposes
Application
SilverLine can be used on the following Dräger de-
vices:
Fabius
Apollo
Babyleo TN500
For further information on system compatibilities,
see the lists of accessories of the main devices or
other documents issued by Dräger.
*
Units of measurement for pressure: 1 kPa x 100 =
100 kPa = 1 bar = 14.5 psi
**
Only for USA
Instructions for use SilverLine
Indications
SilverLine is used to deliver a specific pressure and
flow to the gas inlet of a medical device.
Overview
D
E
B
A
C
G
F
A Main device
B Inlet port, screw connection, e.g., according to
DIN 477-1 or CGA V-1
C Pin-index yoke-type connection according to
EN 850, ISO 407 or CGA V-1
D Pressure gauge
E High-pressure sensor
F Outlet port, NIST connection according to
ISO 18082
G Outlet-port, hose with NIST screw connection
Preparation
WARNING
Risk of user injury
Only use clean, undamaged, reprocessed parts.
 Use dust-free and dry gases only.
 Keep the valve of the compressed gas cylinder
clean.
NOTE
Use the rating plate to check whether the present
pressure reducer is suitable for the intended use
(gas type, pressure, flow). The maximum permissi-
ble input pressure of the pressure reducer must be
equal to or higher than the filling pressure of the
compressed gas cylinder.
1
In well ventilated rooms or outdoors, slowly open
the valve of the compressed gas cylinder for a
short time to blow out any impurities prior to con-
necting the pressure reducer.
WARNING
Risk of injury
Pressure surges induced by the quick opening
of the valve may cause explosions.
Do not direct the gas jet at persons.
WARNING
Risk of material damage and leakage
Tighten connectors only by hand. Using tools
may damage the connectors or cause them to
be over-tightened.
English US
37