Idiomas disponibles

Check weighing
Example: Target weight = 1.000 kg
Switching between checkweighing and normal weighing
➜ Press
to switch between checkweighing and normal weighing.
F
4.4 Classifying
For this purpose, F2.1=OVEr and F2.2.1=CLASS must be set in the operator menu.
In the factory setting, the classifying function is working with upper and lower tolerances of
10 d, 20 d, 30 d. With parameters F2.2.3 and F2.2.4, these tolerances can be customized.
Setting target weight
1. Press
to activate the classifying function.
F
2. Press and hold
If F2.2.2=WEIGHt (factory setting) is set in the operator menu, the weight display appears.
3. Put the target weight on the weighing platform and save with
The OK indicator lights.
If F2.2.2=MAnUAL is set in the operator menu, the weight display with blinking last digit
appears.
3. Enter target weight using the
page 19).
4. Save entered weight value as target weight using the
until tArGEt and the 3 indicators Under, OK and Over appear.
F
T
• Weight is less than the target weight and
below the lower tolerance value.
The Under indicator lights.
• Weight is within the tolerance values.
The OK indicator lights.
• Weight is more than the target weight and
above the upper tolerance value.
The Over indicator lights.
,
and
F
15
.
F
keys and confirm with
key.