Hand-Held Pressure Indicator
CPH6200-S1 / CPH6200-S2
Menu
Param. Setting
►
▲ or ▼
„Menu"
SEt
Set Configuration: Generic settings
ConF
Unit
mbar, bar, ...
SL
oFF / on
Alti
-200 ... +9999 Altitude: Input of altitude above sea-level [m] (only if on)
rAtE
Slo
FASt
P.dEt
t.AVG
1 ... 120
oFF
P.oFF
1 ... 120
oFF
Out
oFF
SEr
dAC
Adr.
01, 11 ... 91
dAC.
CH1, CH2 or
CH DIF
dAC.0
eg. -5.00 ...
+5.00 mbar
dAC.1
eg. -5.00 ...
+5.00 mbar
SEt
Set Calibration: Adjustment of sensors
CAL
OFS.1
Sensordep.,
e.g. -5.00 ...
+5.00 mbar
OFF
SCL.1
eg. -5.00 ...
+5.00 mbar
OFF
OFS.2
Sensordep.,
e.g. -5.00 ...
+5.00 mbar
OFF
SCL.2
eg. -5.00 ...
+5.00 mbar
OFF
WIKA Operating Instructions Hand-Held Pressure Indicator
Meaning
Unit: Unit of display
Sea-level correction: on or off
Rate: Measuring rate (see chapter 2.3)
Slow: Measuring rate (4 Hz filtered, low power consumption)
Fast: Measuring rate, filtered (> 100 Hz)
Peak detection: fast measuring rate, unfildered (> 100 Hz)
Averaging interval in seconds, used by the averaging function off
Averaging function deactivated
Auto Power-Off delay in minutes. Device will automatically switch
itself off if, during this interval, no key is pressed, or there is no
serial communication.
Auto Power-Off function inactive (continuous operation)
Function of the output: No output function, lowest power
consumption
Output is serial interface
Output is analogue output 0 ... 1 V
Communicationsaddress of interface
Choice of the input to be the source for the analogue output (if
Out = dAC)
Enter desired value at which the analogue output potential should
be 0 V (if Out = dAC)
Enter desired value at which the analogue output potential should
be 1 V (if Out = dAC)
The offset of sensor 1 will be displaced by this value to compen-
sate for deviations in the probe or in the measuring device.
Zero displacement inactive (=0.0 °)
The measuring scale of sensor 1 will be changed by this factor
[%] to compensate deviations of temperature probe or measuring
device.
Scale correction factor inactive (=0.000)
The offset of sensor 2 will be displaced by this value to compen-
sate for deviations in the probe or in the measuring device.
Zero dispacement inactive (=0.0°)
The measuring scale of sensor 2 will be changed by this factor
[%] to compensate deviations of temperature probe or measuring
device.
Scale correction factor inactive (=0.000)