3. Specifications / 4. Design and function
3. Specifications
Specifications
GB
Nominal size
Indication accuracy
Scale ranges
Max. working pressure
(static pressure)
Overpressure safety
Ingress protection
Measuring chamber
with pressure
connection (wetted)
For further specifications see WIKA data sheet PM 07.14 and the order
documentation.
4. Design and function
Description
Pressures p
1
by the magnetic piston under pressure (or magnetic piston and separating
diaphragm for model 700.02).
The difference in pressure causes an axial movement (deflection) of the piston
supported by the measuring range spring.
The deflection resulting from this is taken up by a ring magnet found on
the instrument pointer and appropriately displayed. This design combines
the advantages of complete separation of the measuring system and the
indication while eliminating any leakage to the exterior.
For model 700.01 the volume flow from the ⊕ measuring chamber to the ⊖
measuring chamber is minimised by the constructive design and will not interfere
with standard processes.
These instruments are mainly intended for gas/air supply and preparation
applications where no magnetic particles are involved.
8
WIKA operating instructions differential pressure gauge with piston per ATEX
Model 700.01
80
±3 % of full scale value with
increasing differential pressure
0 ... 400 mbar to 0 ... 10 bar
optionally 100, 250 or 400 bar
either side to maximum working pressure
(exception for model 700.02, scale ranges 0 ... 160 mbar and
0 ... 250 mbar: Overpressure safety up to 50 bar)
lP 54 (EN 60529 / lEC 529)
Stainless steel 1.4571, 2 x G ¼ female, on the right and left
side, in-line (EN 837-1 /7.3)
and p
are given in the ⊕ and ⊖ measuring chambers, separated
2
Model 700.02
±5 % of full scale value with
increasing differential pressure
0 ... 160 mbar to 0 ... 2.5 bar
100 bar
(scale ranges 0 ... 160 mbar
and 0 ... 250 mbar: 50 bar)