Occupant Restraints Instruction:
1. The pelvic restraint belt must be worn low across the front
of the pelvis so that the angle of the pelvic belt is within the
preferred zone of 30 to 75 degrees to the horizontal.
A steeper (greater) angle within the preferred zone is desirable
i.e. closer to, but never exceeding 75degrees. (Fig C)
2. The upper torso restraint belt must fit over the shoulder and
across the chest as illustrated Fig d and e
Restraint belts must be adjusted as tightly as possible
consistent with user comfort.
Restraint belt webbing must not be twisted when in use.
The upper torso restraint belt must fit over the shoulder and
across the shoulder as illustrated in Fig D and E.
Fig.C
PREFERED
Fig.D
Fig.A
Fig.E
ZONE
Fig.F
3. The attachment points to the chair are the outer front side
frames just above the castor, and the rear frame. The straps are
fitted into the tie down brackets,
(See Figs G - H)
4. The tie down symbol (Fig F) on the wheelchair frame
indicates the position of the wheelchair restraint straps. The
straps are then tensioned after the front straps have been fitted
to secure the wheelchair.
Fig.B
PELVIC
POSITIONING
BELT
Fig.G
UPPER TORSO
RESTRAINT
BELT
Fig. H