PCE Instruments ScopiX IV Serie Manual De Instrucciones página 94

Tabla de contenido
[SENSe]:BANDwidth
{[1]|2|3|4}[:RESolution]
[SENSe]:FUNCtion
[SENSe]:RANGe
{[1]|2|3|4} :VOLT
[SENSe]:RANGe
{[1]|2|3|4}:AUTO
[SENSe]:RANGe[1]:CAPA
[SENSe]:RANGe[1]:OHM
(Command/Query)
The BAND{[1]|2|3|4} <Bandwidth> command limits the channel bandwidth to the value of the
parameter [5 kHz ; 1,5 MHz ; 15 MHz ; 0 (no bandwidth limit)].
To the question BAND{[1]|2|3|4}?, the instrument returns the value of the filter cut-off frequency
[5 kHz ; 1,5 MHz ; 15 MHz ; 0 (no bandwidth limit)].
(Command/Query)
FUNC<VOLTage|RESistance|CONTinuity|CAPAcitor|DIODe|POWer|POW3a|POW3b|POW3c>
selects the measurement function on channel 1.
To the question FUNC?, the instrument returns the measure function
to channel 1.
POW3a : Three-phase power with two wattmeter method.
POW3b : Three-phase power on a balanced network with neutral.
POW3b : Three-phase power on a balanced network without neutral.
(Command/Query)
The RANG{[1]|2|3|4}:VOLT <range|MAX|MIN|UP|DOWN> command selects the measurement
range to be used in voltmeter mode for the selected channel.
<range> is a value in NRf format, it may be followed or not by a multiple and by the unit.
By default, it is expressed in volt.
To the question RANG{[1]|2|3|4}:VOLT? the instrument returns the value of the measurement
range of the voltmeter for the selected channel.
Response format:
<range><NL>
value in format <NR3>
(Command/Query)
The RANG{[1]|2|3|4}:AUTO <1|0|ON|OFF> command authorizes or prohibits the autoranging
of the selected channel.
ON|1 activates the autoranging.
OFF|0 deactivates this function.
To the question RANG{[1]|2|3|4}:AUTO? the instrument returns the autoranging status for the
selected channel.
(Command/Query)
The RANG:CAPA <range|MAX|MIN|UP|DOWN> command selects the range of measurement
to be used in capacitance mode.
<range> is a value in format NRf, it may be followed or not by a multiple and by the unit.
By default, the value is expressed in Farad.
To the question RANG:CAPA? the instrument returns the range value of the capacitance.
Response format:
<range><NL>
value in format <NR3>
(Command/Query)
The RANG:OHM <range|MAX|MIN|UP|DOWN> command selects the measurement range to
be used in ohmmeter mode.
<range> is a value in format NRf, it may be followed or not by a multiple and by the unit.
By default, it is expressed in Ohm (Ω).
To the question RANG:OHM? the instrument returns the value of the measurement range of
the ohmmeter.
Response format:
<range><NL>
value in format <NR3>