PCE Instruments ScopiX IV Serie Manual De Instrucciones página 91

Tabla de contenido
Programación
MEASure:LOW?
MEASure:MANual:PHASe?
MEASure:MAXimum?
MEASure:MINimum?
MEASure:NWIDth?
MEASure:PDUTycycle?
MEASure:PERiod?
MEASure:PTPeak?
MEASure:PULse:COUNt?
MEASure:PWIDth?
(Query)
To the question MEAS:LOW? <INT{1|2|3|4}> the instrument returns the low level value of the
selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
(Query)
To the question MEAS: MAN:PHAS?, the instrument returns the phase of ϕ-cursor in relation
to cursors 1 and 2. The difference between the cursor 1 and 2 represents 360°. The cursor 1
equal to 0° and the cursor 2, 360°.
Response format:
<measured value><NL>
value in format <NR2> expressed in degree.
(Query)
To the question MEAS:MAX? <INT{1|2|3|4}> the instrument returns the maximum value of the
selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
(Query)
To the question MEAS:MIN? <INT{1|2|3|4}> the instrument returns the value minimum of the
selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
(Query)
To the question MEAS:NWID? <INT{1|2|3|4}> the instrument returns the negatitive pulse width
of the selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in second.
(Query)
To the question MEAS:PDUT? <INT{1|2|3|4}> the instrument returns the duty cycle of the
selected signal.
Response format:
<measured value><NL>
value in format <NR2> expressed in percent.
(Query)
To the question MEAS:PERiod? <INT{1|2|3|4}> the instrument returns the period of the
selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in second.
(Query)
To the question MEAS:PTP? <INT{1|2|3|4}> the instrument returns the peak-to-peak value of
the selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
(Query)
To the question MEAS:PUL:COUN? <INT{1|2|3|4}> the instrument returns the pulse count on
screen of the selected signal.
Response format:
<measured value><NL>
value in format <NR2>.
(Query)
To the question MEAS:PWID? <INT{1|2|3|4}> the instrument returns the positive pulse width
of the selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in second.
91
Tabla de contenido
loading

Tabla de contenido