DISPlay[:WINDow]
:TRACe:Y:LABel{[1]|2|3|4}
DISPlay[:WINDow]
:TRACe:Y[:SCALe]
:PDIVision{[1]|2|3|4}
DISPlay[:WINDow]
:TRACe :Y:SPACing
DISPlay[:WINDow]:CURS
:AUTO:STATe
DISPlay[:WINDow]:CURSor
:TIME{[1]|2|3}:POSition
DISPlay[:WINDow]:CURSor
:STATe
DISPlay[:WINDow]:CURSor
:PHASe:STATe
DISPlay[:WINDow]:CURSor
:REFerence
(Command/Query)
The DISP:TRAC:Y:LAB{[1]|2|3|4} <"label"> command determines the unit of the selected
signal.
The unit is selected among the upper-case letters of the alphabet (A to Z), and is composed of
a name up to 3 letters.
To the question DISP:TRAC:Y:LAB{[1]|2|3|4}?, the instrument returns the unit of the selected
signal.
(Command/Query)
The command DISP:TRAC:Y:PDIV{[1]|2|3|4}<scale|MAX|MIN> command sets the value of
the probe coefficient for the selected signal.
<scale> is a value at NRf format.
To the question DISP:TRAC:Y:PDIV{[1]|2|3|4}?, the instrument returns the value of the probe
coefficient for the selected signal.
(Command/Query)
The DISP:TRAC:Y:SPAC <LOGarithmic|LINear> command specifies the type of scale
applied to the Y-axis.
To the question DISP:TRAC:Y:SPAC?, the instrument returns the type of scale applied to the
Y-axis.
(Command/Query)
The DISP:CURS:AUTO:STAT <1|0|ON|OFF> command activates or inhibits the unattached
cursors.
ON|1 the cursors 1 and 2 move along the reference signal.
OFF|0 the moving of cursors 1 and 2 is free.
To the question DISP:CURS:AUTO:STAT?, the instrument returns the activation status of
unattached cursor mode.
(Command/Query)
The DISP:CURS:TIME{[1]|2|3}:POS <position|MAX|MIN> command
sets the horizontal position of the selected manual cursor.
<position> is a value in format NRf, it may be followed or not by a multiple and the unit.
By default the value is expressed in second.
This command acts on the manual cursors represented on the screen by the X-symbol
accompanied by an index (1, 2 or ϕ).
To the question DISP:CURS:TIME{[1]|2|3}:POS?, the instrument returns the horizontal
position of the selected manual cursor.
Response format:
<measured value><NL>
value in format <NR3> expressed in second.
(Command/Query)
The DISP:CURS:STAT <1|0|ON|OFF> command activates or inhibits the manual
measurements.
1|ON: activates the manual measurements
0|OFF: inhibits the manual measurements
To the question DISP:CURS:STAT?, the instrument returns the activation status of the manual
measurements.
(Command/Query)
The DISP:CURS:PHAS:STAT <1|0|ON|OFF> command activates or inhibits the phase
manual measurement.
To the question DISP:CURS:PHAS:STAT?, the instrument returns the activation status of the
phase manual measurement.
(Command/Query)
The DISP:CURS:REF <INT{1|2|3|4}> command selects the reference for the automatic and
manual measurements.
To the question DISP:CURS:REF?, the instrument returns the signal used as reference.