ARM[:SEQuence{[3]|4}]
:LEVel
AUTOSet:EXEcute
CALCulate:MATH
{[1] |2|3|4}[:EXPRession]
[:DEFine]
CALCulate:MATH
{[1] |2|3|4}[:EXPRession]
:DELete
CALCulate:TRANsform
:FREQuency:WINDow
CALCulate:TRANsform
:FREQuency[:STATe]
DEVice:MODe
DISPlay[:WINDow]
:TRACe:STATe{[1]|2|3|4}
DISPlay[:WINDow]
:TRACe:X[:SCALe]
:PDIVision
(Command/Query)
The ARM:LEV <level|MAX|MIN|UP|DOWN> command sets the trigger level of the auxiliary
source.
<level> is a value in format <NRf>, it may be followed or not by a multiple and by the unit.
By default, the value is expressed in volt.
To the question ARM:LEV?, the instrument returns the trigger level of the auxiliary source.
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
(Command)
The AUTOS:EXE command starts an autoset on each active channel.
(Command/Query)
The CALC:MATH{[1]|2|3|4} <(function)> command defines and activates the mathematical
function of the selected signal.
<function> is the definition of the mathematical function.
(ch1-ch2) subtracts the channel 1 from channel 2.
To the question CALC:MATH{[1]|2|3|4}?, the instrument returns the mathematical function of
the selected signal.
(Command)
The CALC:MATH{[1]|2|3|4}:DEL command deletes the mathematical function of the selected
signal.
(Command/Query)
CALC:TRAN:FREQ:WIND <RECTangular|HAMMing|HANNing|BLACkman|FLATtop>
window used for the FFT calculation.
To the question CALC:TRAN:FREQ:WIND?, the instrument returns the type of window used
for the FFT calculation.
(Command/Query)
The CALC:TRAN:FREQ <1|0|ON|OFF> command activates the FFT calculation.
To the question CALC:TRAN:FREQ?, the instrument returns the activation status of the FFT
calculation.
(Command/Query)
The DEV:MOD <SCOPe|ANALYSer|LOGger|MULTimeter> command selects the principal
mode of the instrument.
To the question DEV:MOD?, the instrument returns the mode in which it has been configured.
(Command/Query)
The DISP:TRAC:STAT{[1]|2|3|4} <1|0|ON|OFF> command validates or devalidates the
selected signal.
To the question DISP:TRAC:STAT{[1]|2|3|4}?, the instrument returns the validation status of
the selected signal.
(Command/Query)
The DISP:TRAC:X:PDIV <scale|MAX|MIN|UP|DOWN> command sets the value of the time
base.
<scale> is a value in format <NRf> , it may be followed or not by a multiple and by the unit.
By default, the value is expressed in second.
Example: to get a time base of 1 µs, following values can be entered: 1E-3ms or 1E-6 or
0.000001s or 0.000001 or else 1us.
To the question DISP:TRAC:X:PDIV?, the instrument returns the value of the time base.
Response format:
<measured value><NL>
value in format <NR3> expressed in second.
86