A 806 Holder For Single-Site® Instruments With Rinse Connection On The Shaft - Miele professional A 600 Instrucciones De Manejo

Módulo inyector
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 55
At the end
of a programme
Fit the stapler into the holder from below.
The pipe connection on the holder must be pushed into the rinse
connection on the stapler.
Position the shaft of the stapler in the V-shaped recess in the sup-
porting rail of the A 502 injector unit.
Align the joint in the functional end of the stapler in the direction of
the water pipe.
Open the functional end of the stapler.
First take the stapler with the hose adapter attached out of the
holder.
Then detach the hose adapter from the shaft of the stapler.
Important. Take care as the hose adapter clamp is very strong.
 The hose adapter must be unscrewed and replaced with a blind
stopper if the holder is not in use. Otherwise the loose end of the
hose adapter will be thrown about by the force of the water and
could cause damage.
en - Areas of application
47