Messages of error are emphasized through an Alfa-numerical code, consisting in a letter
and three numbers.
If an alarm message is diplayed (code "A") the cycle is to be considered
aborted: it will be necessary to repeat all the preparation and sterilization
operations.
To reset alarms and errors hold down at the same time the buttons under the bar indicated
with Reset.
ERROR
A 001
A 101
Vacuum not maintained on the first phase of
A 111
Vacuum not maintained on the second phase
A 121
A132
Pressure above the allowable limit during the
A133
water recall for the warmup phases
A 200
Error in the control of the EV operation
A 400
A 401
A 403
A 405
A 551
A 637
Error in the access to the memory card
Reading of the T1 probe during sterilization
A 651
MESSAGES OF ERRORS
CAUSE
Cycle interrupted by the user
Vacuum not archieved in 10 min.
VACUUM TEST
of VACUUM TEST
Error in the flow-meter operation
Error in the locking operation
Pressure out of limit
above the maximum limit
ENGLISH
SOLUTION
-
Reset and restart the system
-
Check the gasket
-
Check the door adjustment
-
Reset and restart the system
-
Check the gasket
-
Check the door adjustment
-
Reset and restart the system
-
Check the gasket
-
Check the door adjustment
-
Reset and restart the system
-
Reset and restart the system
-
Reset and restart the system
-
Reset and restart the system
-
Reset and restart the system
-
Let cool the room
-
Reset and restart the system
-
check that the SD card is
correctly inserted
-
Reset and restart the system
-
Let cool the room
-
Reset and restart the system
39