Tecno-gaz Multisteril Instrucciones De Uso página 47

Idiomas disponibles
  • ES

Idiomas disponibles

 
9.4
Drain filter cleaning
To be checked periodically.
Make sure no water is present into the tank.
 
9.5
Cleaning of the water infeed filter
To be checked periodically
 
 
The drain filter is located on the rear side of the device
(P).
Unscrew the filter from its housing (the filter might contain
water, before extracting it place a sponge under the filter).
Carefully extract the filter to avoid any damage.
Rinse the filter under tap water to remove impurities.
Clean the filter housing
Carefully fit the cleaned filter into its housing.
Completely screw the filter to avoid water leaks.
 
Close water supply, remove the water drain hose and make
sure the filter (O) is clean.
On the contrary extract it using pliers without forcing and
wash with water.
Remount the filter and the feeding hose.
ENGLISH
19
 

Capítulos

loading