Tecno-gaz Multisteril Instrucciones De Uso página 44

Idiomas disponibles
  • ES

Idiomas disponibles

ENGLISH
 
Disinfectant / detergent infeed
Heating
Degas function
SWEEP function
Ultrasound cleaning
Drying
OK
Press
to scroll the cycle parameters.
SELECT
Press
to modify the value of the selected parameter.
The menu of the parameters and setup scrolls circularly.
Once the setup is completed select EXIT and press
Changes are saved.
7.6
DELAYED START setup:
START
Hold down
SELECT
Press
to scroll values from 1 to 9 hours.
START
Press
to start the delayed start.
During the waiting time a timer is displayed showing the time remaining before the cycle
start.
WARNING: In case voltage is missing the delayed start is cancelled.
8
Open the water feeding cock. Check the presence of the filter in the tank and make sure it
is clean. Position the item to be cleaned into the TRAY in case of small parts use the
DENTAL BURS TRAY.
Items to be cleaned have to be distributed uniformly and must not touch the tank walls or
bottom.
Fit the tray inside the tank, be sure there is no contact between the tank walls and
the tray.
Cements, alginates, and composites must be treated with specific fluids.
In case of rough dirt on the items, such dirt has to be removed before loading the tools
inside the tray to avoid the drain filter obstruction.
To prevent contact between the steel tray feet and the steel cleaning tank rubber rings are
used. In case rings are missing or damaged replace them with new ones. The wrong use
can cause damages to the tank. Switch on the machine through the switch (L) .
Start the selected cycle pressing
the countdown of the time (estimated time) necessary to complete the cycle.
16
 
for 8 seconds.
USE INSTRUCTIONS
START
0 ÷ 6 %
30 ÷ 60 °C
SI - NO
SI - NO
0 ÷ 30 min
SI - NO
SELECT
 
 
the appliance shows the current cycle phase and
1%
10 °C
-
-
1 min
-
to quit the MENU.
LIVEL OF
MAX CAPACITY

Capítulos

loading