Factory Calibration Reset - Lovibond CheckitDirect Manual De Instrucciones

Ocultar thumbs Ver también para CheckitDirect:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Calibration Mode
GB
Store
Date
RESULT
Time
Cal
Zero
Mode
Test
On
Off
SEL
CAL
SEL
cAL
Mode
SEL
CAL
On
Off
40

Factory calibration reset

Resetting the user calibration to the original factory calibration will
reset all methods and ranges.
A user calibrated method is indicated by an arrow while the test result
is displayed.
To reset the calibration press both the [MODE] and [ZERO/TEST] key
and hold.
Switch the unit on using the [ON/OFF] key.
Release the [MODE] and [ZERO/TEST] keys after approx. 1 second.
The following messages will appear in turn on the display:
The factory setting is active.
(SEL stands for Select)
or:
Calibration has been set by the user.
(If the user calibration is to be retained, switch the unit off using the
[ON/OFF] key).
Calibration is reset to the factory setting by pressing the [MODE] key.
The following messages will appear in turn on the display:
Switch the unit off using the [ON/OFF] key.
CheckitDirect_Cl-Br-pH-CyA-TA-CaH_2b 05/2011

Hide quick links:

Tabla de contenido
loading

Este manual también es adecuado para:

Minidirect clMinidirect ph

Tabla de contenido