Functional description

GB
On
Off
CL
Mode
METHOD
Zero
Test
METHOD
0.0.0
Zero
Test
METHOD
RESULT
Zero
Test
Mode
26

Operation

Switch the unit on using the [ON/OFF] key.
The display shows the following:
Select the required test using the [MODE] key:
CL
Br
pH
CyA
The display shows the following:
Fill a clean vial with the water sample up to the 10 ml mark, screw the
cap on and place the vial in the sample chamber making sure that the
marks are aligned.
Press the [ZERO/TEST] key.
The "Method" symbol flashes for approx. 8 seconds.
The display shows the following:
After zero calibration is completed, remove the vial from the sample
chamber. The characteristic coloration appears after the addition of
the reagents.
Replace the cap on the vial and place in the sample chamber making
sure that the
marks are aligned.
Press the [ZERO/TEST] key.
(For Countdown / reaction period see page 27)
The "Method" symbol flashes for approx. 3 seconds.
The result appears in the display.
The result is saved automatically.
Repeating the test:
Press the [ZERO/TEST] key again.
New zero calibration:
Press the [MODE] key until the desired "Method" symbol appears in
the display again.
CheckitDirect_Cl-Br-pH-CyA-TA-CaH_2b 05/2011