Tecno-gaz PURA 10 BIOX Manual Del Usuario página 90

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

ENGLISH
Starting and testing of BIOX: 
A. Fill the glass bottle of Biox with  ClO
(follow instructions in the Biox package). 
B. Fit the bottle into the red cap and fully screw . 
C. Power on the electric component of Biox through the switch  
D. Pick  water  through  the  supplied  devices  or  the  cock  on  the  basin  until 
causing the Biox pump restart. 
E. Hold  down  button  "Manual  dosing"  until  the  dosing  pump  stops  and  the 
"Dosing" yellow LED turns off (approx. 5 seconds). 
F. Switch off Biox and then switch it on. 
G. Repeat three times operations of points E  and F. 
 
H. Hold down button "Filling confirmation". The 6 red LEDs flash for about 5 
seconds, then the LEds turn steady. Release the button. The 30‐day timer 
has been reset . 
I. Continue with the water picking for some dozens of seconds, then wait for 
the machine stop. 
 
Biox is now ready for use. 
 
 
 