This service information describes the connection of the IntelliBridge to the GE Aespire View
anesthesia machine.
Data communication
The IntelliBridge EC10 module supports RS232 serial data communication with the following
specifications for connection with the external device:
• Signaling speed
• Character bits
• Stop bits
RS232 parameters must be set and verified according to the GE Aespire View anesthesia machine
service manual by authorized personnel.
Instruction for connection
Connect the IntelliBridge EC5 D-Sub 15-pin connector to either of the RS232 connector of the GE
Aespire View anesthesia machine as shown below. Make sure to tighten the screws of the connector
properly.
IntelliBridge EC5 connection to RS232 ports
1
2
3
Service Information
7
19200 Bd
7
1
GE Aespire View anesthesia machine - RS232 (COM) ports
IntelliBridge EC5 ID module
EC5 device label: "GE Aespire View"
• Parity generation
• Parity checking
1
2
3
7
Odd
Odd
21