AC measurement (TRMS)
20 A for max. 30 sec. at reduced accuracy
Frequency range 50 Hz...10 kHz
All AC voltage ranges are specified from 5% to 100% of the range
Measurement of AC+DC
Frequency range 0...1000 Hz
20 A for max. 30 sec. at reduced accuracy
Measurement of alternating voltage for frequency >5 kHz
Frequency band 5...100 kHz
Accuracy was determined in the temperature range from 18
below 75%. For distorted current, the error is increased by (2% m.v. + 2% of full scale) for a
crest factor <3.0.
Measurement of resistance
1
When using REL mode to compensate offsets (mode recommended in case of problems with reset-
ting)
Range
Resolution
500.00 μA
0.01 μA
5000.0 μA
0.1 μA
50,000 mA
0.001 mA
500.00 mA
0.01 mA
10,000 A
0.001 A
Range
Resolution
500.00 μA
0.01 μA
5000.0 μA
0.1 μA
50,000 mA
0.001 mA
500.00 mA
0.01 mA
10,000 A
0.001 A
Range
Resolution
50,000 mV
0.001 mV
500.00 mV
0.01 mV
5,0000 V
0,0001 V
50,000 V
0,001 V
Range
Resolution
0.001 Ω
1
50,000Ω
500,00 Ω
0.01 Ω
1
0.0001 kΩ
5,0000 kΩ
0.001 kΩ
50,000 kΩ
0.01 kΩ
500.00 kΩ
0.0001 MΩ
5,0000 MΩ
0.001 MΩ
50,000 MΩ
CMM-60 – USER MANUAL
Basic uncertainty
f = 50/60 Hz
(0.6% m.v. + 25 digits)
f < 1 kHz
(1.5% m.v. + 25 digits)
f < 10 kHz
(3% m.v. + 25 digits)
Basic
uncertainty
(1.0% m.v. + 25 digits)
(1.5% m.v. + 40 digits)
Basic
uncertainty
(5.0% m.v. + 40 digits)
(6.0% m.v. + 40 digits)
Note:
o
C to 28
Basic uncertainty
(0.5 % m.v. + 20 digits)
(0.05 % m.v. + 10 digits)
(0.1 % m.v. + 10 digits)
(0.2 % m.v. + 20 digits)
(2 % m.v. + 20 digits)
o
C and at relative humidity
69