2. Safety / 3. Ex marking
Symbols
EN
3. Ex marking
DANGER!
Danger to life from explosion
Non-observance of these instructions and their contents may result in the
loss of explosion protection.
▶
▶
▶
▶
▶
▶
▶
The precision digital pressure gauge can be operated with substances of
group IIC. When handling the substances, the general rules/provisions of
the substances of group IIC must be observed.
Factors to be tested are:
■
■
■
8
Before mounting and commissioning the instrument, ensure you
read the operating instructions.
Do not dispose of with household waste. Ensure a proper disposal
in accordance with national regulations.
Observe the safety instructions in this chapter and further explosion
instructions in these additional operating instructions.
Observe the information given in the applicable type examination
certificate and the relevant country-specific regulations for installation
and use in hazardous areas (e.g. IEC/EN 60079-14). Follow the
requirements of the ATEX directive.
The precision digital pressure gauge is only suitable for installation
in locations providing adequate protection against intrusion of solid
foreign objects or water which can impair safety.
The use of the precision digital pressure gauge with flammable
substances must be checked by the operator in accordance with the
operating conditions from the specifications.
Temperature range: -10 ... +50 °C [14 ... 122 °F]
Modifications to the instrument will invalidate any approvals.
Only use accessories in hazardous areas that also approved for these
areas.
Media compatibility
Pressure
Temperature
WIKA additional operating instructions, model CPG1500 (Ex i)