6. Safety-related characteristic values
6. Safety-related characteristic values
DANGER!
Danger to life due to loss of explosion protection
The non-observance of the instructions for use in hazardous areas can
lead to the loss of the explosion protection.
▶
Adhere to the following limit values and instructions.
▶
Only use the listed batteries, see chapter 4.2.1 "Approved batteries".
▶
Only replace the batteries outside the hazardous area, see chapter
4.2.2 "Battery replacement".
Permissible temperature ranges
Ambient temperature range (T
Max. medium temperature range (T
Battery supply
Rated voltage (3 x AA-sized batteries)
Max. voltage (spark ignition)
High-frequency output with Bluetooth
Output frequency
Nominal output power
Max. output power (under fault conditions for "ia")
WIKA additional operating instructions, model CPG1500 (Ex i)
)
a
)
m
®
-10 ≤ T
≤ +50 °C
a
-10 ≤ T
≤ +50 °C
m
DC 4.5 V
DC 4.95 V
2,400 ... 2,500 MHz
10 mW
490 mW
EN
13