Test current measurement : Although the current is digitally measured, an
Serial data output
Environmental protection
Safety class
Power supply
Operating temperature
range
Storage temperature range : -25°C to 65°C
Humidity range
Equipment weight
Dimensions
Accessories
Subject to technical change without notice.
analogue taut band instrument shows the result in
order to facilitate the evaluation. This is especially
useful when measuring inductive loads, so that the
operator can verify easily when the test current has
been stabilized.
: RS232 @ 4800 bps. Suitable for data collection in
an external serial printer, computer or data-logger.
: IP54 with closed lid.
: Meets the requirements of IEC 61010-1:1990,
IEC 61010-1:1992 amendment 2
: Mains or internal battery powered
Battery: rechargeable, sealed lead-acid, 12 V -
7 Ah
Internal battery is useful for up to 10 A.
Mains: 220 - 240 V~
Mains power is useful for any test current, including
200 A.
: -5°C to 50°C
: 95% RH (non condensing)
: Approx. 16 kg.
: 502 x 394 x 190 mm
: 2 Combined current and potential leads for 10 A
(1.8 m).
2 Combined current and potential leads for 200 A
(6 m).
1 Communication cable (RS232).
1 Power cord.
1 User guide.
1 Case for the accessories.
18