Connect the Liquid Levels
Sensors to the Hub
Connect each liquid level sensor to the hub port that corresponds
to the reservoir it is mounted on; see Figure 13 and Table 1.
Note: The sensor setup has to be confirmed in the software
when the Biotage
Liquid Level Sensor option is enabled (see
®
"Enable the Liquid Level Sensors" on page 8) and each time
a sensor is connected or disconnected to/from the hub. You can
at anytime see the sensor setup in the About view; see Figure 27
on page 12.
Figure 13. The liquid level sensors connected to the hub.
Connections
Note: The S3 and S4 ports on the hub are not supported on Selekt
Enkel, i.e. sensors connected to these ports will not be used.
System Port
Hub Port
S1
S1
S2
S2
S3
S3
S4
S4
Waste
W
AUX
CAN IN
Table 1. Connections on the system and the liquid level sensor hub.
Use the Universal Cap Adapter
There is an optional universal cap adapter for reservoirs that do
not have the GL45 or 38-430 cap that can be ordered separately
Connect the Liquid Level Sensors to the System
(P/N 418571SP). See the requirements for using the adapter in
Figure 14.
Drill Ø38 mm hole
Min Ø42 mm
Figure 14
. The universal cap adapter can be used with reservoirs that
have a cap with 1) an inner diameter of 42 mm or larger and 2) a maximum
top thickness of 6 mm.
1. Drill a hole of 38 mm in diameter in the middle of the cap;
see Figure 14.
2. Insert the universal cap adapter through the hole; see A and
B in Figure 15.
3. Attach the 38-430 cap supplied by Biotage to the universal
cap adapter; see C in Figure 15.
4. Tighten the 38-430 cap and attach the whole cap assembly
to the reservoir.
A
A
B B
Figure 15
. A: The universal cap adapter. B: The cap with the 38 mm hole
on top of the adapter. C: The 38-430 cap attached to the adapter and the
modified cap.
Figure 16. The liquid level sensor mounted on a reservoir that requires
the universal cap adapter. Do not mount the sensor on a cap that is in
a tilted position.
C C
7