Condition
Hardware Fail
Tubeset Disconnected/
Calibration Required
Description and Corrective Action
This alarm is activated if the pump detects an
internal fault:
• The "spanner" symbol is permanently displayed.
• "ERR" is displayed in the top left of the screen.
• A 2 or 3 digit fault code is displayed in the bottom
right of the screen.
Therapy is suspended.
Switch off the pump.
Call the service engineer.
The fault codes and their descriptions can be
found in the Flowtron ACS900 Service Manual,
part number SER0026.
The Tubeset Disconnected alarm is activated if the
pump detects that the tubeset has been
disconnected from the pump:
• An alarm is activated.
• Both leg icons flash.
• "CAL" is displayed in the top left.
• Therapy is suspended.
The pump must be recalibrated by ArjoHuntleigh
authorised service personnel before therapy can
continue.
Switch off the pump.
Call the service engineer.