Message/symptom
Display: "- - -
No display function
(display OFF)
OFF
Channel cannot be switched
on
No vacuum at desoldering tool − Vacuum not connected
Insufficient vacuum at
desoldering tool
PK Elektronik Vertriebs GmbH, E-Mail: [email protected], Internet: www.pkelektronik.com
9 Error messages and error clearance
Possible cause
− Tool has not been detected
− Tool defective
− No mains supply voltage
− Overload cut-out
− Desoldering nozzle clogged
− Compressed air not or
incorrectly connected
− Filter cartridge on
desoldering tool full
− Main filter on soldering
station full
Remedial measures
− Check connection of tool to
device
− Check connected tool
− Turn on mains power switch
− Check mains supply voltage
− Check device fuse
− Only one soldering iron can be
operated.
− Connect vacuum hose to
vacuum connection
− Service desoldering nozzle
using cleaning tool
− Connect compressed air to
compressed air connection or
check
− Change filter cartridge on
desoldering tool full
− Change the main filter element
on the soldering station (Fig. 41)
Fig. 41