Startup
6.1
Device setup
• Place the Vacuson 40/60 suction pump and all required accessories and instruments on an even, non-slip
surface and make sure you have good access to all controls.
• Do not allow the operating range of the device (including cable) and the connected cnstrument to be com-
promised by limiting factors.
• The manometer must be fully visible at all times.
• The On/Off-AIR-Pedal must be placed within stepping distance between the patient and the surgeon.
• It must be explicitly ensured that no objects can fall on the pedal.
• The power plug at the rear of the device must be accessible at all times.
• The ventilation slots at the housings bottom and sideways of the Vacuson 40/60 must be kept clear in order
to prevent temperature from becoming excessive.
• While in operation the Vacuson 40/60 suction pumps must be at least 1 meter above ground. Nouvag rec-
ommends for convenient handling of the device the use of the optional movable stand.
6.2
Connection to the power supply
Before switching on, make sure that the power supply unit of the device matches the country's specific
service voltage!
The power supply unit of the Vacuson 40/60 pumps is not swichable to the country specific service voltage.
The device has to be ordered according to the country specific service voltage.
In order to prevent the risk of an electric shock, the device may only be connected to a power network
with a PE protective ground conductor.
The power plug socket is located at the rear of the device.
6.3
Preparation of secretion jars
1.
Hold open jar available (2 or 5 liter).
6
2.
Press Jar Lid with Turn and Tilting Lever in open-
position firmly onto the jar (the latch of the lock-
ing system is in open-position).
Turn and Tilting Lever in
open-position.
Operating instructions Vacuson 40/60, Ref. 31997, V43/13
Vacuson 40/60
Latch in open-
position
Gripper
loading