7.2.4 Diagnostic view
The diagnostic view contains important information but is only really useful for troubleshooting
purposes.
7.3 Measurement menu
7.3.1 Instrument configuration
7.3.1.1 Continuous mode description
Continuous mode is typically used for process measurement.
Continuous mode cycle
• Every 2 sec. measurements are refreshed on the display
• The relays and the analog outputs are updated
• Measurements are continuously stored in memory (volatile and non volatile memory) according to
individual settings
7.3.1.2 Continuous mode selection
1. Select Config. instrument from the Measurement menu:
Option
Measurement mode
Pressure
Temperature
7.3.2 Measurement configuration
1. EC sensor
Option
Membrane
Medium
Gas unit type
Gas unit
Liquid
Description
Measurement mode is locked on Continuous
for on line process.
Select the barometric pressure units.
Select the temperature units.
Description
Sensor's membrane number selection.
Liquid or gas phase.
Partial, Fraction, Dissolved.
The list of available units depends on unit type selected above.
Note: This is the gas concentration measured by the EC sensor. When a composite
unit is selected (e.g. ppm » ppb) the unit will change depending on the range of the
value to display.
When medium is liquid, select water or a liquid with a different solubility (if
available).
English 17