Dräger Vista 120 Gateway Manual Del Usuario página 43

Ocultar thumbs Ver también para Vista 120 Gateway:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 289
 Local Listen: local network configuration.
1) IP: Gateway listening IP address,
inbounding data from EMR, default: 0.0.0.0.
2) Device Query Listen: Gateway listening
port, inbounding EMR query request,
default: 9200.
3) ADT Listen: Gateway listening port,
inbounding ADT message from EMR,
default: 9100.
4) Inbound Server IP: the EMR system
allowed to access to Gateway, at most 4
addresses are accepted, default: 127.0.0.1.
 Channel Config: EMR network configuration.
Vista 120 Gateway Instructions for use
1) Channel, IP and Port:
 Channel 1~Channel 4: the target
network configuration for Gateway
outbounding data to EMR, default IP:
127.0.0.1, Port: 3360.
 ADT query: the target network
configuration for Gateway querying ADT
message from EMR, default IP:
127.0.0.1, Port: 3360.
2) Time-out(s): the time that Gateway waits
for response after outbounding data to
EMR, default: 10s.
3) Resend: whether to check resend in case of
time-out, default: unchecked.
Configuration
43
loading