Infusion Times And Flow Rates; Cleaning; Periodic Functional Inspection - Excelsior Medical ESP-60 Manual De Funcionamiento

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Syringe
Volume
Size
5mL
6mL
10mL
12mL
20mL
20mL
30mL
35mL
50mL
60mL
Syringe
Volume
Size
5mL
5mL
10mL
10mL
20mL
20mL
30mL
30mL
50mL
60mL

Cleaning

Clean the exterior surfaces of the unit with a damp cloth and mild detergent or isopropyl alcohol. Avoid getting liquids inside
the device.
CAUTION: Use of harsh cleaning agents or solvents may cause damage to the plastic case and components.

Periodic functional inspection

A periodic functional check should be performed at least every six months. The inspection should include the
following:
1.
Syringe barrel clamp holder should adequately hold each size of syringe (5/6, 10/12, 20, 30/35, 60) firmly in
place.
2.
The pusher arm mechanism should release by pressing the two black release levers and move easily up and down
the unit.
3.
The pusher arm mechanism should engage after releasing the two black release levers.
4.
Press the START NORMAL speed key. The green LED directly above that key should be flashing and motor
operation should be slightly audible.
5.
Place the pusher arm mechanism completely at the bottom of travel next to the syringe barrel clamp holder. The
operation should continue until the end of infusion alarm. Both the flashing red LED and the periodic beep tone
should be operative.
6.
Press the ALARM key. The audible periodic beep tone should cease, but the flashing red LED should continue.
7.
Press the OFF key and then the START NORMAL speed key. The unit should not restart, but continue to indicate
both visual and audible end of syringe alarms.
8.
Repeat steps 4 and 5 for the two other speeds.
9
Pusher arm mechanism and syringe barrel clamp should move freely on the two rails. Use petroleum jelly to
lightly lubricate the two rails and lead screw. DO NOT USE SILICONE LUBRICANTS - DAMAGE TO
PLASTIC MAY OCCUR.

Infusion times and flow rates

(Monoject® Syringes)
Infusion Times
(in minutes)
X-S
Slow Normal
51
34
17
66
44
22
80
53
27
87
58
29
116
78
39
(Becton Dickinson Syringes)
Infusion Times
(in minutes)
X-S
Slow Normal
57
38
19
78
52
26
90
60
30
105
70
35
116
77
39
Flows Rates
mL/hr
X-S
Slow Normal
5.9
8.8
9.2
13.7
15.1
22.6
20.7
31.0
25.8
38.6
Flows Rates
mL/hr
X-S
Slow Normal
5.2
7.8
7.6
11.4
13.3
19.8
17.1
25.6
25.9
38.8
-4-
Nominal
Occlusion
PSI
17.7
41
27.4
26
45.2
16
62.0
12
77.2
9.3
Nominal
Occlusion
PSI
15.7
45
22.7
31
39.7
18
51.1
14
77.5
9.2