4
ENGLISH
2. If you are taking medication, consult with your physician to determine the most appropriate time
to measure your blood pressure. NEVER change a prescribed medication without first consulting
with your physician.
3. For persons with irregular or unstable circulation resulting from diabetes, liver disease,
arteriosclerosis or other medical conditions, there may be variations in blood pressure values
measured at the wrist versus at the upper arm. Monitoring the trends in your blood pressure taken
at either the arm or the wrist is nevertheless useful and important.
4. People suffering from vascular constriction, liver disorders or diabetes, people with cardiac
pacemakers or a weak pulse, and women who are pregnant should consult their physician before
measuring their blood pressure themselves. Different values may be obtained due to their
condition.
5. People suffering from arrhythmias such as atrial or ventricular premature beats or atrial
fibrillation only use this blood pressure monitor in consultation with your doctor. In certain cases
oscillometric measurement method can produce incorrect readings.
6. Too frequent measurements can cause injury to the patient due to blood flow interference.
7.
The cuff should not be applied over a wound as this can cause further injury.
8. DO NOT attach the cuff to a limb being used for IV infusions or any other intravascular access,
therapy or an arterio-venous (A-V) shunt. The cuff inflation can temporarily block blood flow,
potentially causing harm to the patient.
9. The cuff should not be placed on the arm on the side of a mastectomy. In the case of a double
mastectomy use the side of the least dominant arm.
10. Pressurization of the cuff can temporarily cause loss of function of simultaneously used
monitoring equipment on the same limb.
11. A compressed or kinked connection hose may cause continuous cuff pressure resulting in blood
flow interference and potentially harmful injury to the patient.
12. Check that operation of the unit does not result in prolonged impairment of the circulation of the
patient.
13. Product is not intended for infants or individuals who cannot express their intentions.
14. Prolonged over-inflation of the bladder may cause ecchymoma of your arm.
15. Use only the approved arm cuff for this unit. Use of other arm cuffs may result in incorrect
measurement results.
16. The system might produce incorrect readings if stored or used outside the manufacturer's
specified temperature and humidity ranges. Keep the device away from childrens, pets and pests.
17. Advising operator that Instruction manual/ Booklet must be consulted.
18. Do not use the device during patient transport outside healthcare facility for interference source
existing as well.
19. Essential performance:
Electrosurgery interference recovery
Limits of the error of the manometer
Reproducibility of the BLOOD PRESSURE
DETERMINATION
Refer 202.6.2.101 IEC 80601-2-3
Refer 202.12.1.102 IEC 80601-2-30
Refer 201.12.1.107 IEC 80601-2-30