Mounting
Observe the following mounting instructions:
Tilting the sensor can lead to measurement errors.
Always mount the sensor horizontally.
Secure the sensor against movement. Also ensure
that the tank is standing securely.
Make sure that the tank surface and the sensor bot-
tom are clean.
The tank must be easily accessible (min. 80 cm
around the tank).
≥ 20 cm
EN
10
2 m
Mount Sensor
You can mount the sensor with the mounting plate,
straps or your own fixtures.
Notice: Mount sensor horizontal to tank with a maxi-
mum incline of +/- 4°.
Mount Sensor with Mounting Plate
To attach the sensor to the plate, screw the sensor to the
mounting holes provided on both sides using a screw
with a diameter ≤ 5 mm or ≤ M5.
As an alternative to the mounting plate, you can also use
other equipment.
Never drill through the tank! Liquids can leak out.