9. Secure boot straps over each boot foot pad.
NOTE: Pictorial operating instructions are included on the bottom of each boot
(see Figure 2-4).
WARNING - PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:
a. To prevent patient and/or user injury and/or equipment damage, check the
Power-Lift Stirrups and surgical table side rails for possible damage or wear
prior to each use. Do not use the Power-Lift Stirrups if any damage is
apparent, if parts are missing, or if it does not operate as expected. Cracked
or splintered surfaces may cause injury.
b. To prevent patient and/or user injury and/or equipment damage, apply
c. Power-Lift Stirrups only to U.S. standard side rails like those equipped on the
Amsco 3080/3085 SP surgical tables.
d. To prevent patient and/or user injury and/or equipment damage, assure
Document Number: 80028296
Version: A
Figure 2-2. Typical Power-Lift Stirrups Assembly
Figure 2-3. Power-Lift Stirrups Release Handle