Precision of power setting radio frequency (RF)
The power of the radio frequency generator depends on
the tissue specific resistance and can vary within limits.
The specified 50 W are in accordance with a specific
resistance of 1 kΩ. The settings of the radio frequency
will be displayed without the measuring unit [W] and
will be scaled in accordance with the graph.
HF-Leistung in Betriebsart "Cut" vs. Last
RF-power in operating mode „Cut" vs. load
P/W
60
50
40
30
20
10
0
0
500
HF-Leistung vs. Last in Betriebsart "Cut/Coag" bzw. "Coag"
RF-power vs. load in operating mode „Cut/Coag" or „Coag"
bei verschiedenen Leistungseinstellungen, Koagulationsgrad C4
in various power settings, coagulation level C4
P/W
40
30
20
10
0
0
500
HF-Leistung vs. Last in Betriebsart "Bipolar Coag"
RF-power vs. load in operating mode „Bipolar Coag"
bei verschiedenen Leistungseinstellungen, Koagulationsgrad C4
in various power-settings, coagulation level C4
P/W
40
30
20
10
0
0
100
200
300
400
1000
1500
2000
1000
1500
2000
500
600
700
800
900
1000
Max. Ausgangsspannung in Betriebsart "Cut"
Max. output voltage in operating mode „Cut"
U/V
350
300
250
200
25
50
150
100
50
0
0
10
R/W
Max. Ausgangsspannung vs. Anzeige in Betriebsart
Max.output voltage vs. visual display in operating mode
"Cut/Coag" bzw. "Coag", Koagulationsgrad C4
U/V
„Cut/Coag" or „Coag", coagulation level C4
Meyer-Haake Innovations GmbH / Ahlmeier GmbH
350
300
250
200
15
30
150
100
50
0
R/W
0
Max. Ausgangsspannung in Betriebsart
Max. output voltage in operating mode
"Bipolar Coag", Koagulationsgrad C4
U/V
„Bipolar Coag", coagulation level C4
175
150
125
100
15
30
75
50
25
0
R/W
0
User Manual GB
20
30
40
Performance indication
Leistungsanzeige
10
20
Performance indication
Leistungsanzeige
10
20
Leistungsanzeige
Performance indication
50
18.07.2017